The VPFlowScope is available in two flow ranges: 0..80mn/sec and 0..150mn/sec. In the table below you will find common pipe diameters and corresponding flow ranges in m3n/hr and SCFM for both VPFlowScope types.

 

 

 

R080

R150

 

Diameter (mm)

Insertion depth (mm)

Min flow (m3n/hr)
(rounded down to 5 m3n/hr)

Max flow (m3n/hr)
(Rounded down to 5 m3n/hr)

Min flow (m3n/hr)
(rounded down to 5 m3n/hr)

Max flow (m3n/hr)
(Rounded down to 5 m3n/hr)

Measurement uncertainity

25

20

1

135

DO NOT USE

10%

40

35

2

350

DO NOT USE

5%

55

50

4

665

25

1,247

5%

80

40

5

1,445

50

2,705

2%

100

50

10

2,260

80

4,235

2%

160

80

35

5,790

215

10,855

5%

200

100

55

9,045

335

16,955

5%

250

125

85

14,135

530

26,500

5%

 

 

R080

R150

Diameter (mm)

Insertion depth (mm)

Min flow (CFM)
(rounded down to 5 CFM)

Max flow (CFM)
(rounded down to 5 CFM)

Min flow (CFM)
(rounded down to 5 CFM)

Max flow (CFM)
(rounded down to 5 CFM)

Measurement uncertainty

25

20

0.5

80

DO NOT USE

10%

40

35

1

205

DO NOT USE

5%

55

50

2

390

15

735

5%

80

40

3

850

30

1,590

2%

100

50

6

1,330

45

2,490

2%

160

80

20

3,405

125

6,390

5%

200

100

30

5,325

195

9,980

5%

250

125

50

8,320

310

15,600

5%

 

Category